PS1 Off
Power Supply 1 (Short Battery Backup) Not Present / Drive Bus Off
Kuka · Sunrise Cabinet Med Controller
Was bedeutet PS1 Off ?
The PS1 Voltage Green LED is OFF, indicating either that Power Supply 1 (with short battery backup) is not present or that the drive bus is in a 'BusPowerOff' state. This condition could prevent motor drives from receiving power, resulting in a loss of motion control.
Häufige Ursachen
- Power Supply 1 (PS1), potentially for short battery backup, is offline.
- Failure or disconnection of the PS1 unit.
- The drive bus voltage is too low or absent.
- Battery backup for PS1 has failed or is disconnected.
Reparaturschritte & Checkliste
Klicken Sie auf Schritte, um Ihren Fortschritt zu verfolgen.
- 1
Check the infeed at connector X1, verifying the presence and stability of the rated voltage of 27.1 V.