Übersetzen...
PWR/1.2V Off (CIB_SR)

CIB_SR 1.2V Power Supply Not Present

Kuka · Sunrise Cabinet Med Controller

Was bedeutet PWR/1.2V Off (CIB_SR) ?

The PWR/1.2V Green LED for the CIB_SR module is OFF, indicating a lack of 1.2V supply voltage to the module. This absence of critical operating voltage prevents the CIB_SR module from functioning correctly and could result in the non-operation of interconnected I/O nodes.

Häufige Ursachen

  • The 1.2V power supply to the CIB_SR module is not present.
  • The CIB_SR module requires 1.2V power, which is currently absent.
  • Failure or absence of the 1.2V power rail for the CIB_SR communication board.

Reparaturschritte & Checkliste

Klicken Sie auf Schritte, um Ihren Fortschritt zu verfolgen.

  1. 1

    Check fuse F17.3 on the CCU_SR for continuity and proper rating.

  2. 2

    Verify the presence and correct seating of jumper plug X308.

  3. 3

    Check fuse F308 on the CCU_SR for continuity and proper rating.

  4. 4

    If an external supply is used via X308, check the external power supply to ensure it provides the rated voltage of 24 V.

Alle Codes in diesem Handbuch durchsuchen (21)
Verifizierte technische Daten. Zuletzt aktualisiert: March 2026

Verwandte Fehlercodes

Quelle: Kuka Sunrise Cabinet Med Controller
Kuka PWR/1.2V Off (CIB_SR) CIB_SR 1.2V Power Supply Not Pre… | ErrorLib