Übersetzen...
PWRS/3.3V Off
CCU_SR Safety Power Supply Not Present
Kuka · Sunrise Cabinet Med Controller
Was bedeutet PWRS/3.3V Off ?
The PWRS/3.3V Green LED is OFF, indicating an absence of supply voltage for the safety-related 3.3V rail. This condition points to a power supply issue to the CCU_SR module or a fault within the module itself, specifically impacting critical safety functions that rely on this voltage.
Häufige Ursachen
- The safety power supply (PWRS/3.3V) for the safety circuit is not present.
- The 3.3V power rail dedicated to safety functions is off.
- Loss of power to the safety control unit (CCU_SR) safety supply.
Reparaturschritte & Checkliste
Klicken Sie auf Schritte, um Ihren Fortschritt zu verfolgen.
- 1
Check fuse F17.3 on the CCU_SR for continuity and proper rating.
- 2
If the PWR/3.3V LED (Item 6) lights up, indicating the main 3.3V power is present, but the PWRS/3.3V is off, exchange the CCU_SR module as the safety power rail is internally faulty.
Verifizierte technische Daten. Zuletzt aktualisiert: March 2026
Verwandte Fehlercodes
Quelle: Kuka Sunrise Cabinet Med Controller