Übersetzen...
STA2 Off
FPGA Node Power Not Present
Kuka · Sunrise Cabinet Med Controller
Was bedeutet STA2 Off ?
The STA2 Orange LED for the FPGA node is OFF, indicating no supply voltage is present for the FPGA node. This prevents the FPGA node from operating, which is crucial for high-speed logic and data processing within the controller.
Häufige Ursachen
- The FPGA Node (STA2) is not receiving power.
- Main power supply failure affecting the FPGA node.
- Disconnection or damage to the power cable for STA2.
- Issue with the power regulation circuit for STA2.
Reparaturschritte & Checkliste
Klicken Sie auf Schritte, um Ihren Fortschritt zu verfolgen.
- 1
Check the infeed power supply at connector X1.
- 2
If the PWR/3.3V LED (Item 6, Page 98) lights up, indicating the 3.3V power is present, exchange the CCU_SR module, as the fault is likely internal to the module's power distribution for the FPGA node.
Verifizierte technische Daten. Zuletzt aktualisiert: March 2026
Verwandte Fehlercodes
Quelle: Kuka Sunrise Cabinet Med Controller