STA2 Off
FPGA Node Power Not Present
Kuka · Sunrise Cabinet Med Controller
What does STA2 Off mean?
The STA2 Orange LED for the FPGA node is OFF, indicating no supply voltage is present for the FPGA node. This prevents the FPGA node from operating, which is crucial for high-speed logic and data processing within the controller.
Common Causes
- The FPGA Node (STA2) is not receiving power.
- Main power supply failure affecting the FPGA node.
- Disconnection or damage to the power cable for STA2.
- Issue with the power regulation circuit for STA2.
Repair Steps & Checklist
Click steps to track your progress.
- 1
Check the infeed power supply at connector X1.
- 2
If the PWR/3.3V LED (Item 6, Page 98) lights up, indicating the 3.3V power is present, exchange the CCU_SR module, as the fault is likely internal to the module's power distribution for the FPGA node.
Verified technical data. Last updated: March 2026
Related Faults
Source: Kuka Sunrise Cabinet Med Controller