STAS2 Off
Safety Node B Power Supply Not Present
Kuka · Sunrise Cabinet Med Controller
What does STAS2 Off mean?
The STAS2 (Safety node B) Orange LED is OFF, indicating no supply voltage is present for this safety node. This absence of power prevents Safety Node B from operating, which could compromise the overall safety functionality of the system.
Common Causes
- The power supply for Safety Node B (STAS2) is off.
- No power is being supplied to the STAS2 safety device.
- The STAS2 safety module has lost its operational power.
Repair Steps & Checklist
Click steps to track your progress.
- 1
Check fuse F17.3 on the CCU_SR for continuity and proper rating.
- 2
If the PWR/3.3V LED (Item 6, Page 98) lights up, indicating main 3.3V power is present, exchange the CCU_SR module, as the safety node's power supply is likely faulty internally.