Kuka · Sunrise Cabinet Med Controller
The STAS2 (Safety node B) Orange LED is flashing at a rate other than 1 Hz (normal) or 10 Hz (boot phase), indicating an internal fault code within Safety Node B. This suggests a problem with the internal logic, communication, or processing of this safety node, potentially affecting safety chain integrity and preventing safe operation.
Click steps to track your progress.
Check cabling at connectors X309, X310, and X312 for proper connection, integrity, and any signs of damage.
For test purposes, disconnect the cables at X309, X310, and X312, then switch the controller off and back on again to reset the module.
If the error recurs after checking cabling and power cycling, exchange the CCU_SR module.