Temp Fault On

Ballast Resistor R1 Overtemperature

Kuka · Sunrise Cabinet Med Controller

What does Temp Fault On mean?

The 'Temp Fault' Red LED is ON, indicating an overtemperature condition at the ballast resistor R1. This suggests that the ballast resistor is dissipating excessive heat, potentially due to insufficient cooling in the control box, a faulty ballast resistor, or an issue with its temperature sensor. Prolonged overtemperature can lead to component damage or system shutdown.

Common Causes

  • The ballast resistor R1 has reached an overtemperature condition.
  • Excessive heat generated by the ballast resistor during braking or regeneration.
  • The temperature sensor for Ballast Resistor R1 has exceeded its safety limit.

Repair Steps & Checklist

Click steps to track your progress.

  1. 1

    Check the integrity and ensure proper seating of connector X501.

  2. 2

    Inspect the control box cooling system, including fans and vents, for any obstructions or failures.

  3. 3

    Check the temperature sensor located at the central position on ballast resistor R1, which is on the right-hand inner wall of the control box, for proper function and connection.

Browse all codes in this manual (21)
Verified technical data. Last updated: April 2026

Related Faults

27V PS OffMain Power Supply Unit 27V Not PresentFuse DefectiveCabinet Control Unit Fuse DefectivePHY4 OffEthernet Physical Layer Fault
Source: Kuka Sunrise Cabinet Med Controller
Open Manual (Page 101)Search more codes in this manual