KUKA KRC4 · Sunrise Cabinet Med Controller
The STA1 (PMB_SR) LED on the Cabinet Control Unit, Small Robot (CCU_SR) is flashing, indicating an internal fault within the PMB_SR microcontroller responsible for USB functions. This can prevent proper system operation of the KUKA KRC4 Sunrise Cabinet Med Controller.
Click steps to track your progress.
Verify the infeed at connector X1 for the correct supply voltage (rated voltage 27.1 V).
Observe the PWR/5V LED; if it lights up, suggesting 5V power is present, the CCU_SR module may be defective.
Power cycle the KUKA KRC4 Sunrise Cabinet Med Controller to reset the microcontroller.
If the fault persists after initial checks, prepare to exchange the CCU_SR module as indicated.