Traduciendo...
27V PS Off
Main Power Supply Unit 27V Not Present
Kuka · Sunrise Cabinet Med Controller
¿Qué significa 27V PS Off ?
The '27 V Voltage, main power supply unit' Green LED is OFF, indicating that the main power supply unit is not providing its rated voltage. This means the overall system is not receiving the expected 27.1 V input power, which is critical for system operation.
Causas comunes
- The main 27V power supply unit is not providing voltage.
- Internal failure of the 27V power supply.
- Overload on the 27V output causing a shutdown.
- Fault in the input power to the 27V supply.
Pasos de reparación & Lista de verificación
Haga clic en los pasos para seguir su progreso.
- 1
Check the infeed at connector X1, verifying the presence and stability of the rated voltage of 27.1 V.
Datos técnicos verificados. Última actualización: March 2026
Fallas relacionadas
Fuente: Kuka Sunrise Cabinet Med Controller