Traduciendo...
PWR/2.5V Off (CIB_SR)

CIB_SR 2.5V Power Supply Not Present

Kuka · Sunrise Cabinet Med Controller

¿Qué significa PWR/2.5V Off (CIB_SR) ?

The PWR/2.5V Green LED for the CIB_SR module is OFF, indicating a lack of 2.5V supply voltage to the module. This absence of critical operating voltage prevents the CIB_SR module from functioning correctly and could result in the non-operation of interconnected I/O nodes.

Causas comunes

  • The 2.5V power supply to the CIB_SR module is not present.
  • No 2.5V power detected on the CIB_SR board.
  • The CIB_SR communication interface board is missing its 2.5V power input.

Pasos de reparación & Lista de verificación

Haga clic en los pasos para seguir su progreso.

  1. 1

    Check fuse F17.3 on the CCU_SR for continuity and proper rating.

  2. 2

    Verify the presence and correct seating of jumper plug X308.

  3. 3

    Check fuse F308 on the CCU_SR for continuity and proper rating.

  4. 4

    If an external supply is used via X308, check the external power supply to ensure it provides the rated voltage of 24 V.

Ver todos los códigos en este manual (21)
Datos técnicos verificados. Última actualización: March 2026

Fallas relacionadas

Fuente: Kuka Sunrise Cabinet Med Controller