PWR/3.3V Off (CIB_SR)

CIB_SR 3.3V Power Supply Not Present

Kuka · Sunrise Cabinet Med Controller

What does PWR/3.3V Off (CIB_SR) mean?

The PWR/3.3V Green LED for the CIB_SR module is OFF, indicating a lack of 3.3V supply voltage to the module. This absence of critical operating voltage prevents the CIB_SR module from functioning correctly and could result in the non-operation of interconnected I/O nodes.

Repair Steps

  1. 1

    Check fuse F17.3 on the CCU_SR for continuity and proper rating.

  2. 2

    Verify the presence and correct seating of jumper plug X308.

  3. 3

    Check fuse F308 on the CCU_SR for continuity and proper rating.

  4. 4

    If an external supply is used via X308, check the external power supply to ensure it provides the rated voltage of 24 V.

Browse all codes in this manual (21)
Verified technical data. Last updated: March 2026
Source: Kuka Sunrise Cabinet Med Controller