STA2 Flashing
FPGA Node Internal Fault
Kuka · Sunrise Cabinet Med Controller
What does STA2 Flashing mean?
The STA2 Orange LED for the FPGA node is flashing at a rate other than 1 Hz (normal) or 10 Hz (boot phase), indicating an internal fault code within the FPGA node. This suggests a problem with the internal logic or operation of the FPGA, which can severely impact controller performance and stability.
Repair Steps
- 1
Check the infeed power supply at connector X1.
- 2
Verify the status of the PWR/3.3V LED (Item 6, Page 98). If it is lit, exchange the CCU_SR module.
- 3
If the PWR/3.3V LED is not lit, troubleshoot the 3.3V power supply (refer to 'PWR/3.3V Off (CIB_SR)' fault code for specific steps).
- 4
If power supply checks pass and the LED continues to flash, exchange the CCU_SR module.
Verified technical data. Last updated: March 2026
Source: Kuka Sunrise Cabinet Med Controller