STAS2 Flashing

Safety Node B Internal Fault

Kuka · Sunrise Cabinet Med Controller

What does STAS2 Flashing mean?

The STAS2 (Safety node B) Orange LED is flashing at a rate other than 1 Hz (normal) or 10 Hz (boot phase), indicating an internal fault code within Safety Node B. This suggests a problem with the internal logic, communication, or processing of this safety node, potentially affecting safety chain integrity and preventing safe operation.

Repair Steps

  1. 1

    Check cabling at connectors X309, X310, and X312 for proper connection, integrity, and any signs of damage.

  2. 2

    For test purposes, disconnect the cables at X309, X310, and X312, then switch the controller off and back on again to reset the module.

  3. 3

    If the error recurs after checking cabling and power cycling, exchange the CCU_SR module.

Browse all codes in this manual (21)
Verified technical data. Last updated: March 2026
Source: Kuka Sunrise Cabinet Med Controller