Temp Fault On

Ballast Resistor R1 Overtemperature

Kuka · Sunrise Cabinet Med Controller

What does Temp Fault On mean?

The 'Temp Fault' Red LED is ON, indicating an overtemperature condition at the ballast resistor R1. This suggests that the ballast resistor is dissipating excessive heat, potentially due to insufficient cooling in the control box, a faulty ballast resistor, or an issue with its temperature sensor. Prolonged overtemperature can lead to component damage or system shutdown.

Repair Steps

  1. 1

    Check the integrity and ensure proper seating of connector X501.

  2. 2

    Inspect the control box cooling system, including fans and vents, for any obstructions or failures.

  3. 3

    Check the temperature sensor located at the central position on ballast resistor R1, which is on the right-hand inner wall of the control box, for proper function and connection.

Browse all codes in this manual (21)
Verified technical data. Last updated: March 2026
Source: Kuka Sunrise Cabinet Med Controller